Technical management
Traceability
Adjust types, columns, and gaps, then Save. Documents can drop a saved view into a section.
9/11 with V&V
Types
Columns
Status
| Requirement | Parent | Children | Architecture | Design | Risk | V&V | Test coverage | Status |
|---|---|---|---|---|---|---|---|---|
| PN-WSP-SYS-001 SpO2 accuracy | STK-001 | SSY-001 · IF-001 | PMP | SW-CTL · SW-CTL-LP · HW-FLD | — | VER-001 | CHAR-001 · TC-001 · TC-003 · TC-010 | Baselined |
| PN-WSP-SYS-002 Applied-part surface temperature | STK-001 | — | WMR | SW-CTL · SW-CTL-WARM · HW-THM | HAZ-003 · RSK-002 | VER-003 | TC-006 · TC-007 | Baselined |
| PN-WSP-SYS-003 Off-skin and low perfusion blanking | STK-005 | SSY-004 · IF-002 | SNS | SW-PRT · SW-PRT-AIR · SW-ALG-AIR | HAZ-001 | VER-002 | TC-004 · TC-005 | In review |
| PN-WSP-SYS-004 Motion signal quality | STK-005 | SSY-003 | PMP | SW-PRT · SW-PRT-PRES | HAZ-002 | VER-001 | TC-002 | Baselined |
| PN-WSP-SYS-005 Battery endurance | STK-004 | — | BAT | Gap | RSK-002 | VER-004 | — | Approved |
| PN-WSP-SYS-006 Mass | STK-004 | — | ENCL | HW-ENCL | HAZ-007 | VER-004 | — | Approved |
| PN-WSP-SYS-007 Alarm audibility | STK-005 | SSY-002 | HMI | SW-HMI-ALM | HAZ-006 | Gap | — | Baselined |
| PN-WSP-SYS-008 Primary displayed parameters | STK-002 | — | HMI | SW-HMI | HAZ-006 | VAL-001 | TC-009 | Baselined |
| PN-WSP-SYS-009 Basic safety and EMC | STK-005 | — | SYS | HW-PCB | HAZ-005 · RSK-003 | VER-005 | — | In review |
| PN-WSP-SYS-010 Session log | STK-006 | — | MCU | SW-LOG | — | Gap | — | Approved |
| PN-WSP-SYS-011 Time to first valid SpO2 | STK-002 | — | ADH | HW-CAS | HAZ-004 | VAL-001 | TC-008 | Approved |